> @alvinhochun I tested with the FUSB302 not initialized. Is that enough? The FUSB302 only controls CC pins, its entirely separate to all of the QC negotiation over the USB...

This is separate, this is so that the device shows up on USB during normal use (not bootloader mode)

This has been discussed at least twice in the past. Every time it doesn't get anywhere as you can't confidently automatically select the cell count based on voltage alone. Sensing...

It would be possible, though not a trivial amount of work to add. If someone wants to code it into a PR, more than happy to look at it and...

Uhhhh.... That is _very_ weird I have to say. Almost feels like your iron has slower timings than normal. Out of curiosity could you use the debug menu and tell...

A fair bit of head scratching in confusion :confused: . Ill look at it a bit tonight :)

@skoehler Apologies for this getting lost. Is this still occuring for you with latest firmwares?

It should only show that when the tip is disconnected :/ Have not been able to reproduce this at all across an entirely new set of tips.. Out of curiosity...

Your rtip values are reading lower then normal, which most likely means that your iron does not use the exact same op-amp part as other irons, but with calibration _should_...

The calibration of 781 make sense, that is close to your offset. If you power cycle after the calibration is it still showing >500C when the tip is connected?