I tried reading the data from the Hantek 6022BE using libusb1. Oscilloscope with custom firmware (https://github.com/rpcope1/Hantek6022API). Up to 16 MSps on 2 channels and up to 24 MSps on one...

The CY7C68013A reads data from the ADC to the FIFO and transmits data from the FIFO via USB simultaneously. I assume that with a high ADC read rate (samplerate), the...

The actual length of the edges of the calibration output is less than 50 ns (probe = 10MOhm, 15pF). ![scope_325](https://user-images.githubusercontent.com/6552261/32768165-b52844a4-c927-11e7-94d9-e9723f97aaa5.png) ![scope_326](https://user-images.githubusercontent.com/6552261/32768166-b6b82d20-c927-11e7-8a8b-f7286ff7ee8d.png)

The capacitance of the hantek probes stretches the front of the calibration output signal (sorry for my bad english). Yellow - MSO-X probe, green and blue - hantek probes. 1....