Mt Sinai reports:

The device has an automatic shut-off when the pressures exceed 30 cmH2O for >6 sec OR 40 cmH2O for >1 sec. We were able to recreate this by:

  • Setting the device on a high pressure AND
  • Adding in >20LPM O2 AND
  • Almost completely occluding the exhalation port

This limit might live in the configuration section along with the GUI limits, or it might be implemented in code in the therapy manager.