30 cm H2O limit
Mt Sinai reports:
The device has an automatic shut-off when the pressures exceed 30 cmH2O for >6 sec OR 40 cmH2O for >1 sec. We were able to recreate this by:
- Setting the device on a high pressure AND
- Adding in >20LPM O2 AND
- Almost completely occluding the exhalation port
This limit might live in the configuration section along with the GUI limits, or it might be implemented in code in the therapy manager.