David Gault
David Gault
An additional report of this was received in QA-29404 and a sample file provided in QA-29403 The issue was reproduced using Bio-Formats 6.5.1 With that sample file the exposure times...
And additional report of this issue has been received on https://github.com/ome/openmicroscopy/issues/6266 Migrating the original comment from @CFGrote which is below: Summary We experience problems with how exposure time is reported...
Hi @loicsauteur, thank you for the additional details, unfortunately we do not yet have a solution for this particular bug. If you have any sample files showing the issue it...
Thanks @loicsauteur for the sample images, they are a great help. With the latest Bio-Formats (6.6.1) I am seeing the skewing issue pretty clearly in 6 of them, however `HistologzNikonMPVIVO_M2S2-75-25_HE_R2.3.nd2`...
An additional report of this issue has been received on forum thread: https://forum.image.sc/t/problem-with-nd2-large-image-stitch-opening-strange-with-bioformats-on-fiji/67319/5 A sample image showing the skewed image was provided and the error could be reproduced using Bio-Formats...
Hi @antonkulaga, thank you for opening the issue and providing a sample file. I have not been able to reproduce the issue when using Bio-Formats directly. Would you have a...
Thanks @antonkulaga, I will try to reproduce with the samples provided and update this issue once I've learnt more
As a quick update, I can confirm that the issue is reproducible directly with Bio-Formats 6.6.1 using the sample files provided
Some initial debugging suggests that this is occuring when the POIFS tries to read the very first block which is null, the remaining blocks look to be populated and correct....
Hi @yaokkkkkk, if the labels and thumbnail images are included they will appear as additional series on the image. How are you accessing the images? If you are using the...