Orbitrap recognized in `referenceableParamGroupList` but not in `instrumentConfigurationList` ?
This is the file produced by ThermoRawFileParser with default parameters:
<?xml version="1.0" encoding="utf-8"?>
<mzML xmlns:xsi="http://www.w3.org/2001/XMLSchema-instance" xsi:schemaLocation="http://psi.hupo.org/ms/mzml http://psidev.info/files/ms/mzML/xsd/mzML1.1.0.xsd" version="1.1.0" id="060623_std_30sec_CarboAmmo_10mM_70-20-10_01" xmlns="http://psi.hupo.org/ms/mzml">
<cvList count="2">
<cv id="MS" fullName="Mass spectrometry ontology" version="4.1.79" URI="https://raw.githubusercontent.com/HUPO-PSI/psi-ms-CV/master/psi-ms.obo" />
<cv id="UO" fullName="Unit Ontology" version="09:04:2014" URI="https://raw.githubusercontent.com/bio-ontology-research-group/unit-ontology/master/unit.obo" />
</cvList>
<fileDescription>
<fileContent>
<cvParam cvRef="MS" accession="MS:1000579" value="" name="MS1 spectrum" />
<cvParam cvRef="MS" accession="MS:1000810" value="" name="ion current chromatogram" />
</fileContent>
<sourceFileList count="1">
<sourceFile id="RAW1" name="060623_std_30sec_CarboAmmo_10mM_70-20-10_01" location="file:////Users/elliotfontaine/raws/060623_std_30sec_CarboAmmo_10mM_70-20-10_01.raw">
<cvParam cvRef="MS" accession="MS:1000768" value="" name="Thermo nativeID format" />
<cvParam cvRef="MS" accession="MS:1000563" value="" name="Thermo RAW format" />
<cvParam cvRef="MS" accession="MS:1000569" value="2aa446994a290815bb32f4194999d78e62777488" name="SHA-1" />
</sourceFile>
</sourceFileList>
</fileDescription>
<referenceableParamGroupList count="1">
<referenceableParamGroup id="commonInstrumentParams">
<cvParam cvRef="MS" accession="MS:1001742" value="" name="LTQ Orbitrap Velos" />
<cvParam cvRef="MS" accession="MS:1000529" value="SN03173B" name="instrument serial number" />
</referenceableParamGroup>
</referenceableParamGroupList>
<softwareList count="1">
<software id="ThermoRawFileParser" version="1.4.3">
<cvParam cvRef="MS" accession="MS:1003145" name="ThermoRawFileParser" />
</software>
</softwareList>
<instrumentConfigurationList count="1">
<instrumentConfiguration id="IC1">
<referenceableParamGroupRef ref="commonInstrumentParams" />
<componentList count="3">
<source order="1">
<cvParam cvRef="MS" accession="MS:1000073" value="" name="electrospray ionization" />
</source>
<analyzer order="2">
<cvParam cvRef="MS" accession="MS:1000079" value="" name="fourier transform ion cyclotron resonance mass spectrometer" />
</analyzer>
<detector order="3">
<cvParam cvRef="MS" accession="MS:1000624" value="" name="inductive detector" />
</detector>
</componentList>
</instrumentConfiguration>
</instrumentConfigurationList>
<dataProcessingList count="1">
<dataProcessing id="ThermoRawFileParserProcessing">
<processingMethod order="0" softwareRef="ThermoRawFileParser">
<cvParam cvRef="MS" accession="MS:1000544" value="" name="Conversion to mzML" />
</processingMethod>
<processingMethod order="1" softwareRef="ThermoRawFileParser">
<cvParam cvRef="MS" accession="MS:1000035" value="" name="peak picking" />
</processingMethod>
</dataProcessing>
</dataProcessingList>
<run id="_060623_std_30sec_CarboAmmo_10mM_70-20-10_01" defaultInstrumentConfigurationRef="IC1" startTimeStamp="2023-06-21T17:30:53.439Z" defaultSourceFileRef="RAW1">
</run>
</mzML>
I don't think the analyser is a "fourier transform ion cyclotron resonance mass spectrometer". This is the one produced by MSConvert for comparison.
<?xml version="1.0" encoding="utf-8"?>
<indexedmzML xmlns="http://psi.hupo.org/ms/mzml" xmlns:xsi="http://www.w3.org/2001/XMLSchema-instance" xsi:schemaLocation="http://psi.hupo.org/ms/mzml http://psidev.info/files/ms/mzML/xsd/mzML1.1.2_idx.xsd">
<mzML xmlns="http://psi.hupo.org/ms/mzml" xmlns:xsi="http://www.w3.org/2001/XMLSchema-instance" xsi:schemaLocation="http://psi.hupo.org/ms/mzml http://psidev.info/files/ms/mzML/xsd/mzML1.1.0.xsd" id="060623_std_30sec_CarboAmmo_10mM_70-20-10_01" version="1.1.0">
<cvList count="2">
<cv id="MS" fullName="Proteomics Standards Initiative Mass Spectrometry Ontology" version="4.1.56" URI="https://raw.githubusercontent.com/HUPO-PSI/psi-ms-CV/master/psi-ms.obo"/>
<cv id="UO" fullName="Unit Ontology" version="09:04:2014" URI="https://raw.githubusercontent.com/bio-ontology-research-group/unit-ontology/master/unit.obo"/>
</cvList>
<fileDescription>
<fileContent>
<cvParam cvRef="MS" accession="MS:1000579" name="MS1 spectrum" value=""/>
</fileContent>
<sourceFileList count="1">
<sourceFile id="RAW1" name="060623_std_30sec_CarboAmmo_10mM_70-20-10_01.raw" location="file:///T:\redacted\rawdata\230612_Opti-Timepoints">
<cvParam cvRef="MS" accession="MS:1000768" name="Thermo nativeID format" value=""/>
<cvParam cvRef="MS" accession="MS:1000563" name="Thermo RAW format" value=""/>
<cvParam cvRef="MS" accession="MS:1000569" name="SHA-1" value="2aa446994a290815bb32f4194999d78e62777488"/>
</sourceFile>
</sourceFileList>
</fileDescription>
<referenceableParamGroupList count="1">
<referenceableParamGroup id="CommonInstrumentParams">
<cvParam cvRef="MS" accession="MS:1001742" name="LTQ Orbitrap Velos" value=""/>
<cvParam cvRef="MS" accession="MS:1000529" name="instrument serial number" value="SN03173B"/>
</referenceableParamGroup>
</referenceableParamGroupList>
<sampleList count="1">
<sample id="_x0031_" name="">
<cvParam cvRef="MS" accession="MS:1000002" name="sample name" value="1"/>
</sample>
</sampleList>
<softwareList count="2">
<software id="Xcalibur" version="2.7.0 SP1">
<cvParam cvRef="MS" accession="MS:1000532" name="Xcalibur" value=""/>
</software>
<software id="pwiz" version="3.0.22129">
<cvParam cvRef="MS" accession="MS:1000615" name="ProteoWizard software" value=""/>
</software>
</softwareList>
<instrumentConfigurationList count="2">
<instrumentConfiguration id="IC1">
<referenceableParamGroupRef ref="CommonInstrumentParams"/>
<componentList count="3">
<source order="1">
<cvParam cvRef="MS" accession="MS:1000073" name="electrospray ionization" value=""/>
<cvParam cvRef="MS" accession="MS:1000057" name="electrospray inlet" value=""/>
</source>
<analyzer order="2">
<cvParam cvRef="MS" accession="MS:1000484" name="orbitrap" value=""/>
</analyzer>
<detector order="3">
<cvParam cvRef="MS" accession="MS:1000624" name="inductive detector" value=""/>
</detector>
</componentList>
<softwareRef ref="Xcalibur"/>
</instrumentConfiguration>
<instrumentConfiguration id="IC2">
<referenceableParamGroupRef ref="CommonInstrumentParams"/>
<componentList count="3">
<source order="1">
<cvParam cvRef="MS" accession="MS:1000073" name="electrospray ionization" value=""/>
<cvParam cvRef="MS" accession="MS:1000057" name="electrospray inlet" value=""/>
</source>
<analyzer order="2">
<cvParam cvRef="MS" accession="MS:1000083" name="radial ejection linear ion trap" value=""/>
</analyzer>
<detector order="3">
<cvParam cvRef="MS" accession="MS:1000253" name="electron multiplier" value=""/>
</detector>
</componentList>
<softwareRef ref="Xcalibur"/>
</instrumentConfiguration>
</instrumentConfigurationList>
<dataProcessingList count="1">
<dataProcessing id="pwiz_Reader_Thermo_conversion">
<processingMethod order="0" softwareRef="pwiz">
<cvParam cvRef="MS" accession="MS:1000544" name="Conversion to mzML" value=""/>
</processingMethod>
<processingMethod order="1" softwareRef="pwiz">
<cvParam cvRef="MS" accession="MS:1000035" name="peak picking" value=""/>
<userParam name="Thermo/Xcalibur peak picking"/>
</processingMethod>
</dataProcessing>
</dataProcessingList>
<run id="_x0030_60623_std_30sec_CarboAmmo_10mM_70-20-10_01" defaultInstrumentConfigurationRef="IC1" startTimeStamp="2023-06-21T15:30:53Z" defaultSourceFileRef="RAW1">
</run>
</mzML>
<indexList count="2">
</indexList>
<indexListOffset>522043</indexListOffset>
<fileChecksum>822dbcbf725d7759dcd0026dd7d0a6e9061d1d77</fileChecksum>
</indexedmzML>
Hi @elliotfontaine,
thank you for using ThermoRawFileParser. Internally RAW files use the same analyzer type (FTMS) for all FT-enabled mass data (FTICR and Orbitrap in that regard). The <referenceableParamGroup> element contains the instrument model and serial number - these are saved in the RAW file, elements of <componentList> are populated based on the actual scan types used in a particular RAW file. Currently, TRFP assigns MS:1000079 (i.e. FTICR analyzer) for all FTMS scan types. It is possible to deduce if FTMS means FTICR or Orbitrap, based on the instrument model (basically, using a dictionary of all instruments). This feature can be implemented in the next release.