Miloslav Macko
Miloslav Macko
Hi banderlog, is the instrument PR100? Can you post the csv file here? It seems the content has the termination characters used for the message control. Do you need to...
I see the problem... the payload contains the CRLF combinations, but the termination character for transmission is set to LF. The transfer is interrupted by the first LF occurrence, the...
Hi chrisu-mi, sorry for the late answer. What instrument are you communicating with?
Hi Chrisu-mi, with the RTM3004, try this constructor: Instrument = RsInstrument(visaResourceAddr, True, False, options='AddTermCharToWriteBinBlock=True')
I hope that helped, Chrisu-mi, if not, please reopen the issue.
Hi Tomasz, sorry, I overlooked the issue. Is it still actual? Seems like the instrument returns some strange characters. It might help switching to binary transfer, especially if you transfer...
Hi Essence9999, sorry for the delay. Is this issue still active? I see you are using the ESW8. My guess is - the instrument is in a mode where it...
Closed because of the inactivity
Hi, sorry for the late response. If it is still an issue, you can always use raw socket communication for which you do not need any VISA installation: https://rsinstrument.readthedocs.io/en/latest/StepByStepGuide.html#no-visa-session
Hi Bernard-Connektica, sorry I forgot to respond, the Version 1.82.1.106 (13.06.2024) solved this issue.