suspect
suspect copied to clipboard
Siemens DICOM vs rda transform
Should the transform
computed for a Siemens DICOM and the associated exported RDA be exactly the same?
I've got some data where this seems not to be the case, nor do the computed row_vector
and column_vector
match. Unfortunately I can't attach the full data here, but here are some snippets:
relevant part of rda header:
SequenceName: *svs_se
SequenceDescription: *svs_se
[...]
SliceThickness: 15.000000
PositionVector[0]: -33.703157
PositionVector[1]: 21.192990
PositionVector[2]: -27.051218
RowVector[0]: -0.000503
RowVector[1]: -0.099769
RowVector[2]: 0.995010
ColumnVector[0]: -0.003086
ColumnVector[1]: 0.995006
ColumnVector[2]: 0.099767
VOIPositionSag: -33.753210
VOIPositionCor: 35.369816
VOIPositionTra: -18.092139
VOIThickness: 15.000000
VOIPhaseFOV: 30.000000
VOIReadoutFOV: 15.000000
VOINormalSag: 0.999995
VOINormalCor: 0.003020
VOINormalTra: 0.000808
VOIRotationInPlane: -0.099933
FoVHeight: 30.000000
FoVWidth: 15.000000
FoV3D: 15.000000
PercentOfRectFoV: 1.000000
NumberOfRows: 1
NumberOfColumns: 1
NumberOf3DParts: 1
PixelSpacingRow: 30.000000
PixelSpacingCol: 15.000000
PixelSpacing3D: 15.000000
And dumped values from the DICOM:
csa_header["VoiInPlaneRotation"] = -0.099933
csa_header["VoiOrientation"] = [0.999995, 0.00302, 0.000808]
csa_header["PixelSpacing"] = [30.0, 15.0]
csa_header["SliceThickness"] = 15.0
csa_header["VoiPosition"] = [-33.75321, 35.369816, -18.092139]