AndroidAPS
AndroidAPS copied to clipboard
DST: Pump connection lost
Version 2.8.2 Phone: Pixel5 CGM: Dexcom G6 patched app Pump: DanaRS
This year during the time change AAPS issued an alarm exactly at 03:00 due to a possible loss of connection to the pump. The limit entered in the options was 30min. I suspect that here simply the time since the last connection was checked and after the time change of course suddenly a whole hour was in between. My time zone is Europe/Berlin.
Proposal: During the time changeover, the timeout for the connection to the pump is set to 1h15min.
Exactly the same issue here tonight (v3.2.0.3, with Dana I). I hope this can be fixed, as suggested by nshtg.