Hi, I'm an Elekta user and I was wondering if you can incorporate tools for Elekta machine such as DynamicLogFiles, IVIEW EPID, XVI CBCT, Agility MLC...I can provide you data...

**Elekta's EPID images aren't exported with gantry, collimator and couch angles** As the pylinac's WinstonLutz module description states, Elekta's EPID images doesn't contain information about the angles at wich images...

**Describe the bug** Catphan 604 HU Slice not correctly picking center of module **To Reproduce** Steps to reproduce the behavior: 1. How Do I provide you my dataset? 1. Run...

(V2.3.1) Catphan, results() Number of low contrast region "seen" is shown, but test against tolerance (pass/no pass) is not. This value is already available in CTP515 properties (overall_passed), and so...

Context: We implement WL-like test with conical cones collimator (4mm) to check target positionning (by ExacTrac, on a TrueBeam). Method: Step 1: position 4mm BB with exactrac, check BB centering...

(V2.3.1) I generate a PDF report for CatPhan analysis. I would like to include some extra information, regarding acquisition conditions (date/time, device, kV used, ...). We think this can be...

Going forward I plan to use both the Pylinac 2.2.6 Wlutz algorithm and the 3.* Wlutz algorithm in unison. To achieve this it is possible that I can do a...

**Describe the bug** A clear and concise description of what the bug is. The calculation are different on different computers . There is problem in core directory file image.py change...

**Describe the bug** related MLC leaf indexing order issue to [#157](https://github.com/jrkerns/pylinac/issues/157) in picketfence.MLCmeas.leaf_pair function **Expected behavior** In the recent modification of the MLC indexing, the leaf pair is set to...