An easy to read blog about T1 measurements: https://www.nanalysis.com/nmready-blog/part-1-t1-relaxation-definition-measurement-and-practical-implications so the workflow for T1 calculation could be started from pseudo-2D, so nmrium should be able to extract a slice, process...
some tickets are added in order to complete this task: - https://github.com/cheminfo/brukerconverter/issues/122 - https://github.com/cheminfo/nmr-processing/issues/367
On this [blog](https://u-of-o-nmr-facility.blogspot.com/2007/10/t1-measurements-and-estimation.html), seems for 1D T1 measurements (t1ir1d) from bruker (folder experiment) we have `d7` delay constant in seconds. for varian brand (`s2pul` pulse sequence), `d2` delay constant. @lpatiny...
currently for T1 measurement (bruker and jcamp) after split the pseudo2D the `info.vd` contains the information about the delay time or Tau.
Yes we have some code for this. From the zones picked. We could extract the multiplicity

what do you think about split the discontinue spectrum into several continue spectra, in that way nmrium could manage correctly those ROIs
@djeanner, I was wondering about the expected behavior in that case?
I should remark that it does not happen for DIGMOD = 1 (digital). somebody know if we are missing something until to apply the FFT?
At the moment to apply the digital filter by move the group delay from the beginning to the end of the FID, we should take care about an additional zero...