Decouple Rprog setting from internal GND of DCP405
In the current implementation of the DCP405 Rprog setting is done relative to the internal GND of the DCP405. This makes it impossible to connect a AWG with GND to protective earth (PE) AND also connect a scope (with PE to GND) to the - terminal fo the DCP405 due to the voltage difference between the GND and - output of the DCP405.
Attached schematic solves this use case: Both Rprog and the negative potential of the Load can be connected to PE.
- I did not test or design this for cases where PE is connected to the + side of the load (e.q. use the DCP405 to provide negative voltage).
- It looks like there is an 50mV offset in Rprog programming. This probably can be solved with the remark in the schematic. I did not test this.
This issue can be combined with #87?