multigeiger voltage measurements
Parts:
- 2x multigeiger v1.4 board, si22g tube, correct diodes
- 1x multigeiger v1.4 board, lnd712 tube, correct diodes (addtl. 100v zener)
- DVM 10MOhm inner resistance (also measured as 10MOhm)
- 96MOhm (measured, made with serial 9x10MOhm addtl. resistor)
Measurement arrangement:
Vtube
|
96M (serial 9x 10MOhm)
|
10M (DVM inner resistance)
|
GND
Results:
- 371V for Si22G (plateau: 350-475V, recommended 400V)
- 458V for LND712 (plateau: 450-650V, recommended 500V)
Remark: 106MOhm load on the voltage source, thus the measured voltage might be lower than what's there without that addtl. load.
Note: if somebody has a much higher resistance (and still relatively precise) equipment available, it would be interesting to repeat the measurement with that.
@rexfue can you measure this more precisely / with less load?