Open emelalkim opened this issue 4 years ago • 42 comments
trafficstars

I have been working on AIM to DICOM SR conversion. I am using the Cornerstone adapters because we are using Cornerstone and Cornerstone tools for our frontend I have found some missing features and started adding them We worked with David Clunie on AIM DICOM-SR harmonization efforts and I am using that as a reference

Currently I have added

  • comment
  • procedure reported from modality
  • patient info
  • qualitative evaluations

I still have more to implement but I wanted to put this here and ask your opinion/comments Appreciate any input Also Steve said Igor is taking over DICOM SR, what is his github handle? I'd like to ask his opinion too

emelalkim avatar Apr 06 '21 20:04 emelalkim

I also have a question I am not sure who is the best person to ask @JamesAPetts, is that you? So each AIM has a name and Davis Clunie added as the tracking identifier You are using auto generated tracking identifiers, and I saw some checks about valid cornerstore tracking identifier Is it ok if I edit the tracking identifier from the options (like I did trackingUniqueIdentifier)?

emelalkim avatar Apr 06 '21 20:04 emelalkim

This is very timely and we should put our heads together on this. There has been a lot of SR activity also in python for highdicom and also for pathology annotations and AI results, so we should be sure to develop good conventions for interoperability and consistency. @hackermd @cpbridge @seandoyle @fedorov @dclunie maybe we should have a larger group discussion meeting?

@igoroctaviano can you also help review this?

pieper avatar Apr 07 '21 12:04 pieper

@emelalkim could you share some representative image series with the accompanying SRs that demonstrate the capabilities of your implementation?

fedorov avatar Apr 07 '21 13:04 fedorov

@fedorov I put the sample data set I am using and the draft DICOM SR here. I also put some screenshots. I am going to add some tests after I get a more mature conversion. (FYI, we are working on adding radelements so the AIM file has some made up entities for the codes, please ignore them) @pieper I am certainly open to a larger or smaller group meeting. And looking forward to comments from @igoroctaviano and @JamesAPetts

emelalkim avatar Apr 10 '21 01:04 emelalkim

@emelalkim I took a look at your sample, which looks pretty good.

Wrt. TID 1500 compliance:

Content item 1.5.1.4 CONTAINS: CODE: (G-C0E3,SRT,"Finding Site") = (RID58,RadLex,"liver")

should have a "HAS CONCEPT MOD" relationship with its parent (125007,DCM,"Measurement Group"), not "CONTAINS" http://dicom.nema.org/medical/dicom/current/output/chtml/part16/chapter_A.html#para_67e2ac2d-62c4-47d1-851c-975e33b49fed

You are identifying the time point in a non-standard place with a non-standard code:

Wrt. codes:

If you are going to use RADLEX or RADELEMENT, please use the standard all-capitalized coding scheme designators, rather than the lower case ones used now:

dciodvfy reports:

Warning - Unrecognized defined term <RadLex> for value 1 of attribute <Coding Scheme Designator> Warning - Unrecognized defined term <Radelement> for value 1 of attribute <Coding Scheme Designator>

See http://dicom.nema.org/medical/dicom/current/output/chtml/part16/chapter_8.html#para_f568450e-c147-41e3-ba6d-057ffd233e89

Also, it would be better to use SCT rather than SRT codes, since the former are deprecated.

E.g., instead of (G-A185,SRT,"Long Axis"), use (103339001,SCT,"Long Axis")

which is why dciodvfy reports:

Warning - Coding Scheme Designator is deprecated - attribute <CodingSchemeDesignator> = <SRT>

See http://dicom.nema.org/medical/dicom/current/output/chtml/part16/chapter_8.html#sect_8.1

You will probably not be convinced, but in DICOM we prefer to use SCT codes for anatomy, rather than RadLex, i.e., (10200004, SCT, "Liver") rather than (RID58,RADLEX,"liver")

For the finding, you might consider (108369006, SCT, "Tumor") instead of (RECIST_v2,99EPAD,"Tumor assessment").

For specifying it is a target lesion, you might consider (112016, DCM, "Baseline Category") = (112074, DCM, "Target Lesion at Baseline") (see the pattern at http://dicom.nema.org/medical/dicom/current/output/chtml/part16/sect_TID_4102.html#para_929b7c79-ee18-49b2-9eb1-7b7452354fac]

For enhancing lesions, DICOM SEGs use (C113842, NCIt, "Enhancing Lesion") rather than (RID6055,RADLEX,"Enhancing") http://ncit.nci.nih.gov/ncitbrowser/ConceptReport.jsp?dictionary=NCI_Thesaurus&code=C113842

For the lesion "status", for the presence or absence of things, DICOM SR templates usually use the generic SNOMED CT codes, e.g., (52101004, SCT, "Present"); see http://dicom.nema.org/medical/dicom/current/output/chtml/part16/sect_CID_241.html

dclunie avatar Apr 10 '21 13:04 dclunie

Thank you very much @dclunie! This is very helpful. I appreciate you taking the time to go through it. I have put some comments below and have some questions.

@emelalkim I took a look at your sample, which looks pretty good.

Wrt. TID 1500 compliance:

Content item 1.5.1.4 CONTAINS: CODE: (G-C0E3,SRT,"Finding Site") = (RID58,RadLex,"liver")

should have a "HAS CONCEPT MOD" relationship with its parent (125007,DCM,"Measurement Group"), not "CONTAINS" http://dicom.nema.org/medical/dicom/current/output/chtml/part16/chapter_A.html#para_67e2ac2d-62c4-47d1-851c-975e33b49fed

This is coming from dcmjs I think, I will correct it.

You are identifying the time point in a non-standard place with a non-standard code:

  • it would be better to send (126070, DCM, "Subject Time Point Identifier") as a child of (125007,DCM,"Measurement Group") with a TEXT (not NUM) value that identifies the time point, per http://dicom.nema.org/medical/dicom/current/output/chtml/part16/chapter_A.html#sect_TID_1502
  • rather than sending NUM: (RDETBD2,Radelement,"Timepoint") = 0 (linear,UCUM,"linear") in (C0034375,UMLS,"Qualitative Evaluations"); this is also not a valid child of (C0034375,UMLS,"Qualitative Evaluations") since only CODE and TEXT, not NUM are described for that container (per TID 1500 Row 12 http://dicom.nema.org/medical/dicom/current/output/chtml/part16/chapter_A.html#para_e6303ba8-e73d-47e2-b3f3-a8ef763670e2)

OK I'll move it there. Actually we were thinking of moving all the qualitative evaluations under Measurement Group as those are also about the ROI, in a similar manner to quantitative values. What do you think about that? Small question, is the "Subject Time Point Identifier" right term to put? I would think something like a Study Time Point would fit better. Is there a term like that?

  • BTW. "linear" is not a valid UCUM unit - this is moot if you use TEXT rather than NUM, but when a number is dimensionless use "no units"; see http://dicom.nema.org/medical/dicom/current/output/chtml/part16/sect_7.2.2.html and http://dicom.nema.org/medical/dicom/current/output/chtml/part16/chapter_G.html#para_14ea7894-2856-462f-9a62-0ebee7805dff

Good to know I'll use that if I need to put something without units later

  • there is no need to send an Image Library entry for the image if you are not going to say anything about it (provide any content items that describe it)

I think we have the library because the measurements refer to the images. no? I think we had this with the xslt transformation too. Am I wrong?

Wrt. codes:

If you are going to use RADLEX or RADELEMENT, please use the standard all-capitalized coding scheme designators, rather than the lower case ones used now:

dciodvfy reports:

Warning - Unrecognized defined term for value 1 of attribute Warning - Unrecognized defined term for value 1 of attribute

See http://dicom.nema.org/medical/dicom/current/output/chtml/part16/chapter_8.html#para_f568450e-c147-41e3-ba6d-057ffd233e89

Will do

Also, it would be better to use SCT rather than SRT codes, since the former are deprecated.

E.g., instead of (G-A185,SRT,"Long Axis"), use (103339001,SCT,"Long Axis")

which is why dciodvfy reports:

Warning - Coding Scheme Designator is deprecated - attribute =

See http://dicom.nema.org/medical/dicom/current/output/chtml/part16/chapter_8.html#sect_8.1

Makes sense. we also have them in AIM but these are coming from dcmjs. @pieper is it ok for me to change them? I'll fix the AIM ones

You will probably not be convinced, but in DICOM we prefer to use SCT codes for anatomy, rather than RadLex, i.e., (10200004, SCT, "Liver") rather than (RID58,RADLEX,"liver")

These are coming from the RECIST template we have in ePAD. I'll talk with Daniel on how we can change it, for old AIMs especially

For the finding, you might consider (108369006, SCT, "Tumor") instead of (RECIST_v2,99EPAD,"Tumor assessment").

This is the template's controlled term saying this is a RECIST template. finding was the place you put the template coded term when we were doing the xslt transformation. Do you think it is a good idea to just put Tumor? I'll discuss with Daniel also

For specifying it is a target lesion, you might consider (112016, DCM, "Baseline Category") = (112074, DCM, "Target Lesion at Baseline") (see the pattern at http://dicom.nema.org/medical/dicom/current/output/chtml/part16/sect_TID_4102.html#para_929b7c79-ee18-49b2-9eb1-7b7452354fac]

How important do you think this is? Because I'm not sure this is possible. In the RECIST template in ePAD, time point and lesion type are defined separately. I would have to alter the data/terms when converting to DICOM SR. which is doable but very specific and I think it would be very confusing for the user

For enhancing lesions, DICOM SEGs use (C113842, NCIt, "Enhancing Lesion") rather than (RID6055,RADLEX,"Enhancing") http://ncit.nci.nih.gov/ncitbrowser/ConceptReport.jsp?dictionary=NCI_Thesaurus&code=C113842

For the lesion "status", for the presence or absence of things, DICOM SR templates usually use the generic SNOMED CT codes, e.g., (52101004, SCT, "Present"); see http://dicom.nema.org/medical/dicom/current/output/chtml/part16/sect_CID_241.html

There 2 are also coming from the template. If we end up changing the template, which seems like we will, I'll fix it all

emelalkim avatar Apr 10 '21 20:04 emelalkim

Also, it would be better to use SCT rather than SRT codes, since the former are deprecated.

E.g., instead of (G-A185,SRT,"Long Axis"), use (103339001,SCT,"Long Axis")

which is why dciodvfy reports:

Warning - Coding Scheme Designator is deprecated - attribute =

See http://dicom.nema.org/medical/dicom/current/output/chtml/part16/chapter_8.html#sect_8.1

Makes sense. we also have them in AIM but these are coming from dcmjs. @pieper is it ok for me to change them? I'll fix the AIM ones

Yes, agreed, it would be great if you can update the coding in dcmjs.

Great to see progress on this @emelalkim and thank you @dclunie for your help.

pieper avatar Apr 14 '21 13:04 pieper

I have just one comments. "Tracking Identifier" is populated to what looks like a mapping to the application-specific tool used to do the annotation. I believe the intent for "Tracking identifier" is to be a human-readable identifier of the finding to facilitate tracking of the finding over time.

fedorov avatar Apr 14 '21 19:04 fedorov

Also, it would be better to use SCT rather than SRT codes, since the former are deprecated. E.g., instead of (G-A185,SRT,"Long Axis"), use (103339001,SCT,"Long Axis") which is why dciodvfy reports: Warning - Coding Scheme Designator is deprecated - attribute = See http://dicom.nema.org/medical/dicom/current/output/chtml/part16/chapter_8.html#sect_8.1

Makes sense. we also have them in AIM but these are coming from dcmjs. @pieper is it ok for me to change them? I'll fix the AIM ones

Yes, agreed, it would be great if you can update the coding in dcmjs.

Great to see progress on this @emelalkim and thank you @dclunie for your help.

Will do @pieper

emelalkim avatar Apr 14 '21 20:04 emelalkim

I have just one comments. "Tracking Identifier" is populated to what looks like a mapping to the application-specific tool used to do the annotation. I believe the intent for "Tracking identifier" is to be a human-readable identifier of the finding to facilitate tracking of the finding over time.

Thanks @fedorov, I'll put the annotation name. I just wanted to have @JamesAPetts's opinion first to see if it will mess up cornerstone as it seems to be doing some checks. I don't understand that part. I didn't hear back yet. If I don't hear back in a couple of days. I'll change it and try to see if it will mess up loading in cornerstone

emelalkim avatar Apr 14 '21 20:04 emelalkim

I just wanted to have @JamesAPetts's opinion first to see if it will mess up cornerstone as it seems to be doing some checks.

If it does - we should revisit this as part of the IDC work. I don't think it should!

fedorov avatar Apr 14 '21 20:04 fedorov

I just wanted to have @JamesAPetts's opinion first to see if it will mess up cornerstone as it seems to be doing some checks.

If it does - we should revisit this as part of the IDC work. I don't think it should!

ok. I'll proceed with adding annotation name then. Thanks

emelalkim avatar Apr 14 '21 20:04 emelalkim

I have just one comments. "Tracking Identifier" is populated to what looks like a mapping to the application-specific tool used to do the annotation. I believe the intent for "Tracking identifier" is to be a human-readable identifier of the finding to facilitate tracking of the finding over time.

@fedorov - I don't want to hijack this thread - but I was wondering if the "Tracking Identifier" was meant to be immutable? Here's my use case: in the lumbar spine SR that I'm making I create both a tracking identifier and a tracking uid for labeling each lumbar spine disc (there is also a separate SNOMED code for the disc). But a user can shift the vertebral labels up and down if they disagree with the original labeling. I was planning in this case to change the "Tracking Identifier" but not the uid. We want to use the uid to track the changes that the user made.

An alternative would be to just use a tracking identifier of 'disc1', 'disc2' and make them immutable. This would avoid the problem of the user changing labels with the result being that they were inconsistent.

seandoyle avatar Apr 15 '21 14:04 seandoyle

I have just one comments. "Tracking Identifier" is populated to what looks like a mapping to the application-specific tool used to do the annotation. I believe the intent for "Tracking identifier" is to be a human-readable identifier of the finding to facilitate tracking of the finding over time.

Thanks @fedorov, I'll put the annotation name. I just wanted to have @JamesAPetts's opinion first to see if it will mess up cornerstone as it seems to be doing some checks. I don't understand that part. I didn't hear back yet. If I don't hear back in a couple of days. I'll change it and try to see if it will mess up loading in cornerstone

Yes, the current implementation uses a tracking identifier as a key to map the measurement group to its cornerstone annotation tool-type equivalent. In this issue: https://github.com/OHIF/Viewers/issues/1215 we are discussing new alternatives of doing this because storing vendor-specific names in tracking identifiers seems too brittle and we should use a smart matching strategy or something along those lines e.g. bidirectional measurement properties matched? (e.g. match created based on Clunie's planner annotations paper) then assign the correct vendor tool type.

Example of planar annotation stowed using OHIF:

CodeMeaning - Tracking Identifier
TextValue - cornerstoneTools@^4.0.0:Bidirectional
Tracking Unique Identifier (112040 - DCM)
RelationshipType - HAS OBS CONTEXT

where the first bit of the string identifies a vendor, the second bit is the version, and the last bit the vendor annotation tool type. This string should be interpreted and given to MeasurementService in order to map the SR content to its corresponding annotation format using a mapper implemented by the vendor and registered in OHIF. This is not done currently since this service is still in progress, so what's happening is that inside OHIF the SR is interpreted directly and parsed into annotations. More about this service here: https://github.com/dannyrb/Viewers/blob/f99a6699107c932fb067168ef13ea787d00855c8/docs/latest/services/data/measurements.md

The measurement service + mappings ideal + ideal way of storing things is still an open discussion.

Your incremental changes don't seem to affect our STOWRS/SR logic for cornerstone annotations. You're adding a comment property to bidirectional (which I think is not interpreted by cstools) and changing tracking unique identifiers which I think is currently ignored when generating the cornerstone report in DCMJS. @JamesAPetts feel free to correct me :)

igoroctaviano avatar Apr 15 '21 21:04 igoroctaviano

@igoroctaviano, that doesn't sound like it is going to work if you receive annotations in DICOM SR from another application that doesn't follow your magic convention for how to fill in Tracking Identifier. You should map based on the form of the annotation and allow the Tracking Identifier to be whatever it needs to be for the use case per local procedures (e.g., in a previous life, we required these to follow a certain convention per SOP, such as NT002 for non-target lesion #2, to match entries in our clinical database management system, and not to be constrained by the image annotation tool).

dclunie avatar Apr 16 '21 10:04 dclunie

@seandoyle, nothing should be immutable in the workflow (as opposed to a single serialized instance of an SR) - errors happen, so when a user-mislabeled lesion needs to be renamed (e.g., NT001 to NT002), it might (or might not) be appropriate to keep the same UID, depending on the workflow and local convention (e.g., what needs to happen when a new/updated SR with the corrected information is saved and ingested by the downstream systems that use the information, e.g., RECIST calculations).

In your intervertebral disk labeling use case, I might ask the questions:

"what does the user want to see?" to which the answer would seem to be "L2/3" not "disc A" or similar, corrected on a subsequent occasion or not, and

"does it matter if the UID changes" to which the answer is "it depends", on the impact on the downstream workflow, e.g, are there post-processing measurements/decisions indexed by that UID stored separately, such as "thickness", that should remain associated with that annotation even if its human-readable identifier has changed, or those that should not, such as semantic information such as (67459009, SCT, "Intervertebral disc, L2-L3")

Note that in tracking corrections like this, DICOM SR has a predecessor reference mechanism that can be used to determine the head (most recent) "version" to be used (and thence to suppress earlier versions from use), if changes need to be tracked.

dclunie avatar Apr 16 '21 10:04 dclunie

@igoroctaviano, that doesn't sound like it is going to work if you receive annotations in DICOM SR from another application that doesn't follow your magic convention for how to fill in Tracking Identifier. You should map based on the form of the annotation and allow the Tracking Identifier to be whatever it needs to be for the use case per local procedures (e.g., in a previous life, we required these to follow a certain convention per SOP, such as NT002 for non-target lesion #2, to match entries in our clinical database management system, and not to be constrained by the image annotation tool).

Exactly, that's is the current OHIF approach which at the time was done just as a POC to demo stowrs I guess. This map you mentioned is what I meant in the last comment ...we should use a smart matching strategy or something along those lines e.g. bidirectional measurement properties matched? then assign the correct vendor tool type (e.g. match criteria created based on Clunie's planner annotations paper) . That instead of the current "magic"

I think it's not so easy to come with solid matchers for all tools but we can start with simple ones and then incrementing them as we go. I'll try working on at least one e2e example using measurement service + matching criteria + no tracking identifier and then we can get the ball rolling

igoroctaviano avatar Apr 16 '21 11:04 igoroctaviano

Hi @igoroctaviano and @dclunie, How about inferring from the points in the SR. I mean SR already has the graphic types of POINT, MULTIPOINT, POLYLINE, CIRCLE and ELLIPSE. The problematic one is POLYLINE, which can be length, bidirectional or freehand tool in cornerstone. and with bidirectional there are 2 lines. and you can compute if they are perpendicular. that is how we do it in AIM. Another question is how to handle multiple shapes in one SR. I mean is it possible for the SR to include measurements for two separate lines or two freehand shapes for example? I assume it is, how to handle it then?

emelalkim avatar Apr 21 '21 02:04 emelalkim

Also @dclunie, I talked with Daniel and I am going to update the template. Can you clarify these for me? -we were thinking of moving all the qualitative evaluations under Measurement Group as those are also about the ROI, in a similar manner to quantitative values. What do you think about that? Small question, is the "Subject Time Point Identifier" right term to put? I would think something like a Study Time Point would fit better. Is there a term like that? -I think we have the image library because the measurements refer to the images. no? I think we had this with the xslt transformation too. Am I wrong?

  • About RECIST_v2 coded term :This is the template's controlled term saying this is a RECIST template. finding was the place you put the template coded term when we were doing the xslt transformation. Do you think it is a good idea to just put Tumor?
  • Baseline=Target lesion suggestion: How important do you think this is? Because I'm not sure this is possible. In the RECIST template in ePAD, time point and lesion type are defined separately. I would have to alter the data/terms when converting to DICOM SR. which is doable but very specific and I think it would be very confusing for the user

emelalkim avatar Apr 21 '21 02:04 emelalkim

Hi, @igoroctaviano, @pieper I am working on writing DICOM SRs for segmentation AIMs. The segmentation objects in adapters are not classes like the other tools. Is there any implementation about that I didn't see? should I convert the segmentation export to a class so that it can be used with MeasurementReport?

emelalkim avatar May 21 '21 06:05 emelalkim

Hi Emel -

I'm not following your question exactly. The way we've been handling things is that the SEG contains the pixel data of the segmentation (and things like anatomic codes and color) and the measurements are in the SR that references the SEG. So if you have an SR that says the tumor is 3cc, it references the SEG which references, say, the PET it was defined on. That's how we have it this data: https://wiki.cancerimagingarchive.net/display/Public/QIN-HEADNECK

The adaptors in dcmjs are meant to map between the dicom instances and the in-memory representations for the various libraries. So the concept of a segmentation as it exists in OHIF and/or ePad probably maps to both a SEG and SR where the measurement in SR references the SEG. I don't recall that we've addressed that case yet in dcmjs, but it is in slicer/dcmqi and I believe now it should also exist in highdicom (maybe @hackermd can comment on that?). In any case we should aim for interoperability and similar patterns across implementations.

pieper avatar May 21 '21 19:05 pieper

Hi @pieper, sorry I wasn't clear. I couldn't see any code in these adapters supporting writing dicom sr for segmentations/referring them. does OHIF support that? if yes, how?

emelalkim avatar May 21 '21 20:05 emelalkim

writing dicom sr for segmentations/referring them. does OHIF support that?

I'm pretty sure nobody has done that yet. What I would do is look at the structure of the QIN-HEADNECK data I linked above and just copy it into a new dataset object for conversion to part10. Then as far as I know the best way to test it would be to load it into Slicer's dicom database with the QuantitativeReporting extension installed. If everything is referenced the same as the QIN-HEADNECK conventions then doubleclicking the SR should offer to load the SEG and the referenced images as well. I have done that before for SEG from OHIF, and SR from CrowdsCure, but not SR that references SEG.

pieper avatar May 21 '21 20:05 pieper

@pieper @emelalkim the classes in dcmjs.sr.templates should support that in principle. These are pretty much a 1-1 translation of the highdicom implementation. However, not all parts may have been extensively tested so far on the JS side.

hackermd avatar May 24 '21 17:05 hackermd

As we work on this can we add some tests for the sr code just to show how all the pieces should fit together?

pieper avatar May 24 '21 22:05 pieper

Hi @igoroctaviano and @dclunie, How about inferring from the points in the SR. I mean SR already has the graphic types of POINT, MULTIPOINT, POLYLINE, CIRCLE and ELLIPSE. The problematic one is POLYLINE, which can be length, bidirectional or freehand tool in cornerstone. and with bidirectional there are 2 lines. and you can compute if they are perpendicular. that is how we do it in AIM. Another question is how to handle multiple shapes in one SR. I mean is it possible for the SR to include measurements for two separate lines or two freehand shapes for example? I assume it is, how to handle it then?

Hey @emelalkim, I was rereading the issue and I think I missed this comment. As Danny said here, I think that's the case for related measurements that are linked by UID? (and span measurement groups)? some measurements will explicitly map to a POINT or POLYLINE, while others are implied or derived from neighboring info. Maybe @dclunie can shed some light on it?

Regarding the "in-progress" adapters code I mentioned earlier, just figured that it's not related to freehand or rect (me and my big mouth), it's just a very early adapters code which is basically what we have in dcmjs but scattered. Feel free to implement the freehand and rect adapters, I'll be happy to help to bring people to validate what you have in mind since this is something that's gonna be needed in OHIF as we add freehand and rect SR support. I can even try to reach out to @JamesAPetts or other team members that worked with the adapters before if needed.

igoroctaviano avatar Jun 30 '21 22:06 igoroctaviano

Hi @igoroctaviano and @dclunie, How about inferring from the points in the SR. I mean SR already has the graphic types of POINT, MULTIPOINT, POLYLINE, CIRCLE and ELLIPSE. The problematic one is POLYLINE, which can be length, bidirectional or freehand tool in cornerstone. and with bidirectional there are 2 lines. and you can compute if they are perpendicular. that is how we do it in AIM. Another question is how to handle multiple shapes in one SR. I mean is it possible for the SR to include measurements for two separate lines or two freehand shapes for example? I assume it is, how to handle it then?

Hey @emelalkim, I was rereading the issue and I think I missed this comment. As Danny said here, I think that's the case for related measurements that are linked by UID? (and span measurement groups)? some measurements will explicitly map to a POINT or POLYLINE, while others are implied or derived from neighboring info. Maybe @dclunie can shed some light on it?

Regarding the "in-progress" adapters code I mentioned earlier, just figured that it's not related to freehand or rect (me and my big mouth), it's just a very early adapters code which is basically what we have in dcmjs but scattered. Feel free to implement the freehand and rect adapters, I'll be happy to help to bring people to validate what you have in mind since this is something that's gonna be needed in OHIF as we add freehand and rect SR support. I can even try to reach out to @JamesAPetts or other team members that worked with the adapters before if needed.

Hi @igoroctaviano, do you mean the tracking Unique Identifier? So if there are more than one ROI in one SR, they would have different tracking unique identifiers. About '"in-progress" adapters code I mentioned earlier', you mean there is no code for planar annotations, right?

emelalkim avatar Jun 30 '21 23:06 emelalkim

Hi @igoroctaviano and @dclunie, How about inferring from the points in the SR. I mean SR already has the graphic types of POINT, MULTIPOINT, POLYLINE, CIRCLE and ELLIPSE. The problematic one is POLYLINE, which can be length, bidirectional or freehand tool in cornerstone. and with bidirectional there are 2 lines. and you can compute if they are perpendicular. that is how we do it in AIM. Another question is how to handle multiple shapes in one SR. I mean is it possible for the SR to include measurements for two separate lines or two freehand shapes for example? I assume it is, how to handle it then?

Hey @emelalkim, I was rereading the issue and I think I missed this comment. As Danny said here, I think that's the case for related measurements that are linked by UID? (and span measurement groups)? some measurements will explicitly map to a POINT or POLYLINE, while others are implied or derived from neighboring info. Maybe @dclunie can shed some light on it? Regarding the "in-progress" adapters code I mentioned earlier, just figured that it's not related to freehand or rect (me and my big mouth), it's just a very early adapters code which is basically what we have in dcmjs but scattered. Feel free to implement the freehand and rect adapters, I'll be happy to help to bring people to validate what you have in mind since this is something that's gonna be needed in OHIF as we add freehand and rect SR support. I can even try to reach out to @JamesAPetts or other team members that worked with the adapters before if needed.

Hi @igoroctaviano, do you mean the tracking Unique Identifier? So if there are more than one ROI in one SR, they would have different tracking unique identifiers. About '"in-progress" adapters code I mentioned earlier', you mean there is no code for planar annotations, right?

That's right. 👍

igoroctaviano avatar Jul 01 '21 14:07 igoroctaviano

Note that tracking UID is optional.

fedorov avatar Jul 01 '21 14:07 fedorov

Here is an example in the standard of using the same Tracking Unique Identifier in two different Measurement Groups to indicate that different kinds of measurements apply to the same finding. The example uses a SEG reference to define the ROI but the same approach would apply if one used a closed POLYLINE to define a contour around the lesion instead, since it may still be necessary to use separate Measurement Groups if one is defining an Area or Volume, and the other linear distance measurements.

http://dicom.nema.org/medical/dicom/current/output/chtml/part17/sect_RRR.5.html

Note also that for the bi-dimensional measurements, each of the two perpendicular lines is (a) encoded separately and (b) labelled with a specific code (long vs. short axis).

See also the discussion in:

https://docs.google.com/document/d/1bR6m7foTCzofoZKeIRN5YreBrkjgMcBfNA7r9wXEGR4/edit#heading=h.ag0f07ftbosu

dclunie avatar Jul 01 '21 16:07 dclunie