so if we transmit at say 10 MHz, we may adjust the receive filtering from 15MHz in the near field, and linearly slide it to the imaging depth to say...

if you have specific presets and probe models, we can give you that info

Transmit of 14MHz Receive from 15MHz to 10MHz over 4cm RF does not go through demodulation, therefore no RX applied

thanks for the graph and data, we're not too sure, as our acoustic measurements show probes range from 9.5-10.5mhz center frequency.

you can now change some low level parameters through the cast interface, documentation here: https://github.com/clariusdev/raw#low-level-parameters

Are there any OpenGL errors that get output on the console/debugger? For raw data, you must enable from the App first (assuming research package was purchased) thanks

raw data should have images, but still strange you would not see the real-time image, still could be udp port blocked if you do not see other errors

this one might be different since the callback is actually engaged, but a black image. It may have something to do with the GL rendering, are there any errors in...

Seems that the GPU drivers may be the culprit if you can check out [this case](https://github.com/clariusdev/cast/issues/102) and see if it will help.