AAPS v3.1.0.2 no longer syncs pump manual bolus history for IOB calculations
-
Issue started only after v3.1.0.2
-
has happened twice already
-
mobile switched off. Latest issue at 3:20 pm bolus 1 unit directly on combo pump as shown below

-
Mobile switched on at 6 pm. 1u not used in IOB calculations or treatment tab

-
Please note 1u does showup in combo tab last bolus field but not used elsewhere

-
Not sure if limited to combo pump or all pumps
This should be taken care of in the new combov2 driver, since the command-mode history delta will contain the events that happened since the last time the delta was retrieved. In this case, it should contain an entry that shows that a bolus was manually requested. The history delta has a limited space for entries, though normally, this limit should not be reached (you'd have to deliver lots of manual boluses in a short amount of time for example).